Customization: | Available |
---|---|
Application: | Industry, School, Hospital, Lab |
Customized: | Customized |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
CY-H1200-1L | ||
1 | Tube | Stainless steel tube |
Tube size:OD 60mmx Length 600 (mm) | ||
2 | Working Pressure vs. Working Temperature |
200 Bar max. at 650°C 140 Bar max. at 800°C 80 Bar max. at 900°C 40 Bar max. at 1000°C 2.5 Bar max. at 1100°C Never heat up above 1200°C |
3 | Pressure Measurement Range | 0~15Mpa |
4 | Max. Heating Temperature without Loading |
1100°C |
5 | Working Gases | Inert gas, hydrogen and oxygen |
6 | Heating Temperature | 1100°C max.for < 1hr |
1000C for continuous | ||
7 | Heating Rate | ≤20°C/min |
8 | Heating Element | Fe-Cr-Al Alloy doped by Mo |
9 | Heating Zone | 200mm |
10 | Temp. Control | PID control method with over-temp protection Temperature stability: ±1°C Touch screen controller with LCD display |
11 | Positive Pressure Sensor | Installed in high pressure flange |
12 | PC control | Yes |
13 | Pressure Alarm System | Upper limit and lower limit control method |
Alarm strobe and warning whistle | ||
14 | Voltage | AC 220V 50/60Hz |
15 | Power | 6 KW |
(20A breaker required) | ||
16 | Flanges & Fitting | One pair of CF-63 flange and copper gasket is included |
One 1.5meter long, 1/4" high pressure flexible tube is included for connecting to gas tank and gas-inlet furnace tube. Max. working pressure 200bar. |
||
Please never use ordinary SS tube to connect to gas-inlet for using > 5Mp pressure, which may cause tube exploded. The safety pressure relief valve is available to be installed at the downstream side. |
||
17 | Application Notes | The heating temperature must be limited by the certain pressure level due to vessel material strength. |
The pressure is increasing while the temperature is rising during operation. So you need to calculate the pressure changing towards your target temperature and set a safe value of pressure during process. |
||
18 | Warranty | One year limited manufacturer warranty with lifetime support (Consumable parts such as processing tubes, o-rings and halogen Lamps are not covered by the warranty). |
19 | Pressure safety device | 1)high pressure sensor installed in high-pressure flange 2)high pressure solenoid valve controlled by pressure monitor 3)automatic gas release port 4)pressure transducer 5)manual valve for disconnect solenoid |
20 | Standard Accessories | high pressure super-alloy furnace tube, 1ea high pressure sealing flanges and the related accessories like copper O-rings, screws, 1set high pressure solenoid valve and pressure transducer, 1set dust free protective gloves, 1 pair tube blocks, 1 pair furnace hook, 1ea quartz boat, 1ea |