Working Pressure vs. Working Temperature |
Pressure calculation based on safety factor 3 for 6000 hours service life o 200 Bar max. at ≤800°C o 150 Bar max. at ≤900°C o 100 Bar max. at ≤1000°C o 50 Bar max. at ≤1100°C |
Working Gases | Inert gas and oxygen No hydrogen or any flammable gas |
Heating Temp. | 1100°C max.for < 1hr 1000°C for continuous |
Heating Rate | ≤10°C/min |
Heating Element | Fe-Cr-Al Alloy doped by Mo |
Heating Zone | Two heating zones: 200mm + 200mm Total heating zone length: 400mm Constant temp. zone: 175 mm ±2°C |
Temp. Control |
Three precision digital temperature controller with 30 segments programmable PID control method with protection of over-temp. and thermal couple broken Temperature stability: ±1°C PC control software and interface are included for PC operation |
Positive Pressure Sensor |
One digital pressure transducer installed in high pressure flange One pressure controller is included. the release valve will automatically open when pressure reach setting pressure. Pressure Measurement Range: 0~15Mpa |
Process Vessel | Nickel based super-alloy GH747 (Click to download GH747 nickel based alloy data sheet) Φ70 OD × Φ35 ID ×526 L mm, Normal Capacity: 500 ml Attention: Processing Tube is consumable part, you can click picture below to order replacement or spare one. Optional torque wrench is available here to avoid over-tighten the flange bolts. |
Flange and Fitting |
A pair of 63mm CF type flanges is included (w/ high pressure valve and 1/4 tube fitting) 2 pcs Copper o-rings are included ( spare copper O-ring is available as consumable. The flange is cooling by forced air via aluminum heat-sink |
Optional Parts ( with extra cost ) |
You may order Heavy Duty Rotary Vane Vacuum Pump (7.8 CFM -240 L/m) - EQ-YTP-50 for experiment under vacuum condition Please consider the Turbo-molecular Vacuum Pump Station (up to 1E-7 mbar) - EQ-PV-HVS2 for high vacuum level Gas flow control system is available upon request ( see picture below right ) |
Voltage | AC 208 - 240V 50/60Hz |
Power | 3 KW (20A breaker required) |