Descriptions of vacuum glove box:
CY-VGB-3 is a middle size heavy duty stainless steel glove box with Airlock Chamber, Gauges and vacuum Flanges, which allow glove box operate under vacuum to remove moisture and oxygen without expensive inert gas purging, It is an ideal tool for material and chemical researchers to process air sensitive materials and solutions.
Technical Parameters:
Glove Box Chamber
|
- Case Material: Made of 11 gauge (3mm thickness) 304 stainless steel sheet with support frame and plastic painting, which provide heavy duty structure for vacuum
- Dimensions: 780mm(L) x 700mm(W) x 650mm(H)
- Max. Positive Pressure: 810 Torr (1.1 atm)
- Note: Gloves cannot work if pressure is higher than 810 Torr
- Max. Vacuum Level: 0.5 Torr
- Note: Stainless Steel case may deform at higher vacuum level
|
Airlock Chamber
|
- Dimensions: 240mm(ID) x 260mm(L)
- Max. Vacuum Level: 0.05 Torr
|
Vacuum gauge and valves |
- Two vacuum flanges are included for glove box to achieve vacuum to remove moisture and oxygen without expensive inert gas purging
- Do not operate under vacuum status. The vacuum process is used for purging the glove box chamber before an operation.
- Two KF-25 vacuum fitting ports with Quick Clamps are installed in the back for connecting vacuum pump, electric/fluid feed-through, gas circulating/purifying system, or other testing equipment
|
Leakage Rate in Glove Box Chamber |
< 6 Torr/hour |
Gloves
|
- One pair of Latex Gloves is included for immediate use
- Clamps and sealing flanges included
- Optional: You may order gloves because they have:
- Increased stretching, bending ability and tear strength
- High effectiveness when vaporized hydrogen peroxide (VHP) is present
- High resistance to aromatics, hydrocarbons, solvents, fats, and acids Resists ozone
- High static dissipative qualities
|
Power Extension |
- One extension socket (Max power 1500W) is installed in the glove box main chamber.
- Note: 110V power cord was included for immediate use. (For use under 208~240V power supply, no plug will be included)
|
Vacuum pump and fittings (Optional) |
- Vacuum pump and fittings are not included.
- We may add a customized flange to connect a tube furnace to the glove box
- You may order pressure controller to keep inert gas pressure constant
- You may order the glove box with drying system.
|
Warranty |
- One year limited warranty with lifetime support
- Rusting and damage due to improper storage condition or maintenance are not covered by warranty
- Gloves are consumable items and are not covered under warranty
|
Shipping Dimensions |
50" x 40" x 37" |
Shipping Weight |
240 kg (520 lbs) |
Application Notes:
- Corrosive liquid (such as LiPF6 electrolyte) must be remain sealed in a container inside the glove box.
- Otherwise, liquid vapor may condense and corrode the steel chamber and purification pipeline.
- The H2O & O2 removing rate is largely dependent on the type of purging gas used. For faster chamber purging process,
- Nitrogen is preferred over Argon due to its lighter mass.
- By closing the vacuum flange, the glove box can be vacuumed by mechanical pump up to 0.5 torrs before filling inert gas,
- (moisture can be below 180 ppm and oxygen below 200 ppm). Then you can save at least 1-2 tank of inert gas from purging at each operation
- Please call or email us to get the price and lead time for multi-pins feed through on the glove box if you need to measure the sample inside the chamber,
- CY's technician can provide a solution with no impact on sealing (leaking rate less than 6 torrs/hour) in an affordable price. Please click the picture below to see our perfect customization:
- You may choose humidity and oxygen analyzer below at extra cost to monitor gas humidity and oxygen content inside the glove box
- The humidity inside the glove box can be controlled by using either manual purging system or Gas Purification System. Please see our experimental results below:
- Humidity Range 4~5% Rh: Can be achieved after manually purging the system for 6 cycles by using N2 gas supply tank
- Humidity Range < 2% Rh: Please use either gas supply with 99.9999% gas purity with moisture filter or a Recirculating Gas Purification System for purging
Note: For Li-ion battery research, please order CY-VGB-5, CY-VGB-6, or CY- VGB-7 to achieve humidity < 10ppm
Glove box chamber size:
Air lock and chamber tray:
Optional equipment for your choice:
CE certification of laboratory Vacuum Glove Box:
Contact Information: